Overview

Product name: EasyStep Human LpPLA2(Phospholipase A2, Lipoprotein Associated) ELISA Kit
Reactivity: Human
Alternative Names: PLA2G7; PAF-AH; PAFAH; Lp-PLA2; LDL-PLA2; Platelet Activating Factor Acetylhydrolase,Plasma; Phospholipase A2,Group VII; LDL-associated phospholipase A2
Assay Type: Sandwich
Sensitivity: 0.78 ng/mL
Standard: 50 ng/mL
Detection Range: 0.78-50 ng/mL
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Enzyme & Kinase;Metabolic pathway;Infection immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human LpPLA2, and the Human LpPLA2 standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human LpPLA2. After TMB substrate solution is added, only those wells that contain Human LpPLA2 and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human LpPLA2 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
50.00 1.996 1.910
25.00 1.502 1.416
12.50 1.157 1.071
6.25 0.896 0.810
3.13 0.527 0.441
1.57 0.312 0.226
0.79 0.190 0.104
0.00 0.086 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant LpPLA2 and the recovery rates were calculated by comparing the measured value to the expected amount of LpPLA2 in samples.
Matrix Recovery range Average
serum(n=5) 85-95% 90%
EDTA plasma(n=5) 83-97% 90%
Heparin plasma(n=5) 81-95% 88%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LpPLA2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 90-99% 82-101% 93-105%
EDTA plasma(n=5) 85-100% 84-98% 97-101% 88-96%
Heparin plasma(n=5) 86-93% 91-96% 87-98% 95-102%

产品成分表